Standard 5 m* * On request can other cable construction and length are supplied.
Element type
Ag, AgCl(s), Cl – , Ag +
Purity silver
99.90%
Concentration Chloride ions
0,5 M KCl
Potential
250 mV @25°C relative to. Standard Hydrogen Electrode (SHE)
Accuracy
+/- 5mV
Drift
+/- 1mV in 24 hour @ 5μA load
Temperature coefficient
-0.65mV/°C
Temperature range
-5 to 70°C
Polarisation properties
2 mV possible shift after application or 0,1 µA for 30 seconds in NaOH solution 12 mV possible shift after application or 1 µA for 30 seconds in NaOH solution
Theoretical service life
20 years at 0.1 μA load *The functional service life is in the practice likely determined through the used cabling.
Cable thickness
2,5mm²
Diameter 16mm (+/-1mm)
Length 84mm (+/-2mm)
Diameter Measuring surface • 15mm (+/-1mm)
Cable insulation • XLPE/PVC* Blue
Cable length • Standard 5m*
* On request can other cable construction and length are supplied.
Element type • Ag, AgCl(s), Cl – , Ag +
Purity silver • 99.90%
Concentration Chloride ions • 0,5 M KCl
Potential • 250 mV @25°C relative to. Standard Hydrogen Electrode (SHE)
Accuracy • +/- 5mV
Drift • +/- 1mV in 24 hour @ 5μA load
Temperature coefficient • -0.65mV/°C
Temperature range • -5 to 70°C
Polarisation properties • 2 mV possible shift after application or 0,1 µA for 30 seconds in NaOH solution 12 mV possible shift after application or 1 µA for 30 seconds in NaOH solution
Theoretical service life • 20 years at 0.1 μA load
*The functional service life is in the practice likely determined through the used cabling.